Mbar torr hpa – VACUUBRAND MV 10C VARIO User Manual
Page 22

page 22 of 51
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999073 / 15/12/2009
How to change the pressure units
+
mbar
Torr
hPa
(mains switch)
Keep key p
▲ or p▼ pressed during switching on.
☞ The pressure units are displayed, the pressure unit as from
last operation is flashing.
➨ Press key p▲ or p▼ to change the pressure unit. Press key
START/STOP when the controller displays the desired pres-
sure unit to finish the operation mode.
p
If the automatic mode has been preselected ”End” is displayed.
☞ Switching between automatic switching off activated or not
activated is possible by pressing key
▲ or ▼.
➨ Confirm by pressing key START/STOP.
☞ Automatic switching off as soon as the controller detects
that evaporation is completely finished. The pump is switched
off, the cooling water is running on, if a coolant valve is
preselected.
☞ Without automatic switching off, the pressure in the vacuum
system is adjusted after the evaporation is completed if nec-
essary until the ultimate vacuum of the system.
The controller switches to normal operation mode (process control not active).