How to change the pressure units

+

mbar
Torr
hPa

(mains switch)

Keep key p

▲ or p▼ pressed during switching on.

☞ The pressure units are displayed, the pressure unit as from

last operation is flashing.

➨ Press key p▲ or p▼ to change the pressure unit. Press key

START/STOP when the controller displays the desired pres-
sure unit to finish the operation mode.

p

If the automatic mode has been preselected ”End” is displayed.
☞ Switching between automatic switching off activated or not

activated is possible by pressing key

▲ or ▼.

➨ Confirm by pressing key START/STOP.

☞ Automatic switching off as soon as the controller detects

that evaporation is completely finished. The pump is switched
off, the cooling water is running on, if a coolant valve is
preselected.

☞ Without automatic switching off, the pressure in the vacuum

system is adjusted after the evaporation is completed if nec-
essary until the ultimate vacuum of the system.

The controller switches to normal operation mode (process control not active).