VACUU•LAN:
☞ Operation of the pump depending on actual pressure de-
mand.
☞ The cooling water is switched off if no more pumping is re-
quired, it is switched on again if gas or vapour occur.
☞ The basic setup for decentralized vacuum network systems
(for vacuum in laboratories, vacuum lines for weekend, night
or trainee operations). Continuous vacuum without continu-
ous pumping.
☞ Additional extension of the lifetime of diaphragms and valves
(only as many diaphragm strokes as necessary).
Changing the basic setup
Starting the program:
➨ Keep key MODE pressed while switching on the controller.
☞ The basic setup ”vacuum control” (factory-set) is displayed
on the LCD by flashing symbols.
To change the basic setup:
➨ Press key ▲ or ▼ until the symbol of the desired mode is
flashing.
➨ Confirm by pressing START/STOP.
Only in basic setup vacuum control:
☞ ”Set”, ”Auto” and ”yes” or ”no” are displayed.
➨ Switching between ”yes” or ”no” is possible by pressing key
▲ or ▼.
➨ To preselect the automatic mode confirm ”yes” by pressing
key START/STOP.
➨ If you do not wish to preselect the automatic mode confirm
”no” by pressing key START/STOP.
vacuum control
continuous pumping VACUU•LAN
☞ The symbol of the venting valve is displayed and ”yes” or
”no”.
➨ Switching between ”yes” or ”no” is possible by pressing key
▲ or ▼.
➨ Confirm the selection ”yes” or ”no” by pressing key START/
STOP.
After selecting the basic setup:
☞ The symbol of the coolant valve is displayed and ”yes” or
”no”.
➨ Switching between ”yes” or ”no” is possible by pressing key
▲ or ▼.
➨ Confirm the selection ”yes” or ”no” by pressing key START/
STOP.
MODE
+
(mains switch)