FUNCTIONS OF KEYS, CONTROLS, INDICATIONS, AND CONNECTORS

(cont.)


REAR PANEL

9 pin D-sub connector

Connects to external actuation device, and enable power and frequency
monitoring.

Footswitch jack

Connects to the footswitch cable.

Coax connector

Connects to the converter.

Power module

Connects to the electrical line cord and encases the fuse(s).


9-PIN D-SUB CONNECTOR

Pin No.

Description

1 Not

connected

2 Not

connected

3 Not

connected

4

Enables connection to a frequency counter.

5

Enables connection to an external power monitor (5 mv = 1 watt)

6 Ground
7

Energizes the ultrasonics when connected to ground.

8 and 9

Enables the intensity to be remotely adjusted using an external 10k
potentiometer. See below






NOTE

To vary the intensity remotely using a variable DC power supply (0-5V) instead of a 10 K
potentiometer, connect positive to pin 8 and negative to pin 6.