Set ON/OFF power switch to ON. The switch will illuminate and the screen will
display the power rating of the Ultrasonic Processor, cautionary notices, and the
following control parameters.
CAUTION
•
Never allow liquid to spill into the converter. Do not use the cup horn without a
splash shield
•
Do not allow a microtip or extender to vibrate in air for more than 10 seconds.
When working with a microtip
never
allow the AMPLITUDE control to be set
above the microtip limit 40%. Ignoring these instructions will cause the microtip
to fracture.
•
Do not allow the vibrating microtip to contact anything but the sample.
•
When working with low surface tension liquids, do not use a probe with a
replaceable tip.
•
Never energize a threaded probe without the replaceable tip, extender, or microtip
attached.
NOTE
Refer to Section IV, for general operating suggestions and ultrasonic processing
techniques.
TIME __:__:__ TEMP __ __
°C
PULSE__:__:__ AMPL __ __%