beautypg.com

Sonics VCX750 (Serial No."X or lower)" User Manual

Page 16

background image

16

PULSER:

By inhibiting heat build-up in the sample, the pulse function enables

safe treatment of temperature sensitive samples at high intensity. In addition,
pulsing enhances processing by allowing the material to settle back under the
probe after each burst. The ON and OFF pulse duration can be set independently
from .1 second to 9.9 seconds. During the OFF portion of the cycle, the red
indicator on the

PULSER

key will illuminate. If the OFF portion of the cycle

exceeds two seconds, a cautionary message

- CAUTION – PROBE ON

STANDBY –

will warn the operator against touching the ultrasonic probe. To set

the pulser, press

PULSER

key.



The screen will display:



Using the numeric keys, set the ON portion of the cycle, and press the

ENTER/REVIEW

key.

The screen will display:



Using the numeric keys set the OFF portion of the cycle.

The screen will display:


Press the

ENTER/REVIEW

key.


The screen will display:


Pulse on __.__sec

Pulse off __.__sec

Pulse on 2.5 sec

Pulse off __.__sec

e.g.

Pulse on 2.5 sec

Pulse off 1.0 sec

e.g.

TIME 5:30:25 TEMP __ __

°C

PULSE 2.5 : 1.0 AMPL 40 %

Go To Top Of Document

This manual is related to the following products: