Return of equipment, Temperatur probe calibration, Return of equipment temperature probe calibration – Sonics VCX750 (Serial No."X or lower)" User Manual
Page 22
22
RETURN OF EQUIPMENT
It is suggested that an Ultrasonic Processor in need of repair be sent back to the factory.
In order to receive prompt service; always contact the factory before returning any
instrument. Include date of purchase, model number and serial number. For instruments
not covered by the warranty, a purchase order should be forwarded to avoid unnecessary
delay. Care should be exercised to provide adequate packing to insure against possible
damage in shipment. The Ultrasonic Processor should be sent to the “Service
Department” with all transportation charges prepaid and return of shipment indicated.
Please obtain a Return Authorization Number prior to returning the instrument.
TEMPERATURE PROBE CALIBRATION
Ultrasonic Processors shipped with the optional Temperature Probe have been calibrated
as a set. If the Temperature Probe is acquired separately, calibration should be performed
in accordance with the procedure outlined below.
To calibrate the Ultrasonic Processor, proceed as follows:
1.
Fill a 500 ml vessel with approximately 50% ice and 50% water. Allow 5
minutes for the water temperature to stabilize.
2.
Fill another 500ml vessel with boiling water, and maintain boiling condition
with an immersion heater or other heating device.
3.
Forcefully insert the Temperature Probe into the small jack on the rear panel.
4.
While holding the
TEMP
key depress, set the ON/OFF power switch to ON.
The switch will illuminate and the screen will display the following message:
NOTE
For optimum accuracy, the Temperature Probe and ultrasonic
Processor should be calibrated as a set.
IMPORTANT
I CERTIFY THAT THE ULTRASONIC PROCESSOR AND / OR ACCESSORIES RETURNED
FOR REPAIR ARE FREE OF ANY BIOHAZARDOUS OR RADIOACTIVE MATERIAL AND ARE
SAFE FOR HANDLING.
DO NOT RETURN ANY EQUIPMENT UNLESS SUCH CERTICATION CAN BE MADE.