beautypg.com

Sonics VCX750 (Serial No."X or lower)" User Manual

Page 17

background image

17

TEMPERATURE:

The temperature function prevents overheating of the sample

by continuously monitoring the sample temperature, and terminating the
ultrasonics when the temperature reaches a predetermined setpoint. The
ultrasonics is automatically reinstated when the temperature drops below the
setpoint. If the temperature of the sample must be monitored and /or controlled,
insert the optional Temperature Probe forcefully into the small jack on the rear
panel, immerse the Temperature Probe in the sample and Press the

TEMP

key.

The screen will display:

Using the numeric keys set the high temperature limit (setpoint).

The screen will display:


Press the

ENTER/REVIEW

key.


The screen will display:


REVIEW:

The REVIEW function provides a “window” on the process by displaying

various operating parameters without process interruption. Pressing the

ENTER/REVIEW

key repeatedly during processing will consecutively display the

following information.

a)

Selected amplitude:

e.g. Amplitude Control 40%

b)

Set and read temperature:

e.g. Temp Set 35

°C Probe 27°C

c)

Set processing time and elapsed processing time:

e.g. Set 5:30:25 Time 0:57:03

d)

Selected pulsing cycle, and actual pulsing cycle:

e.g. Pulse 2.5 1.0/1.5 .5

e)

Amount of power in watts, and accumulated amount of energy in JOULES delivered to the
probe:

e.g. 20 watts 0000000 Joules

f)

Elapsed time since processing was initiated:

e.g. Elapsed time 1:27:33

Probe Temperature 27

° C

Temperature Setpoint __

° C

e.g.

Probe Temperature 27

° C

Temperature Setpoint 35

° C

e.g.

TIME 5:30:25 TEMP 35

° C

PULSE 2.5 : 1.0 AMPL 40 %

Go To Top Of Document

This manual is related to the following products: