Culligan Gold Series User Manual
Page 15

• Align the manifold with the center opening in the valve, and press the valve onto the adapter
firmly.
• Assemble the tank clamp to the control, and tighten the clamp screw.
Flow Control Eductor Nozzle
Listed below is the recommended eductor nozzle to be used at various salt dosages.
Table 2 - Flow Restrictors
Unit
Backwash Flow
Nozzle
Throat
Brine Refill Flow
9”
2.0 gpm (#2 Brown)
Blue*
Light Brown*
0.45 gpm
10”
2.0 gpm (#2 Brown)
Beige
Light Brown
0.45 gpm
12”
3.5 gpm (#3 Green)
Beige
Light Brown
0.8 gpm
14”
5.5 gpm (Black)
Green
Blue
0.8 gpm
* Standard from factory
Refer to Figure 6 for a visual on changing the
eductor nozzle and the backwash flow control.
Eductor Nozzle Replacement
• Remove the three screws on the eductor cap
and remove the cap.
• Remove the eductor assembly.
• Remove the eductor screen from the assembly
• Remove the blue nozzle and replace it with the
beige nozzle. Make sure to put the o-ring on the
correct nozzle.
• Reverse the procedure to reassemble. To prevent
leaks, ensure that the gasket is in the proper
position.
Backwash Flow Control Replacement
• Remove the drain clip and pull the drain elbow straight off.
• Remove the backwash flow control located behind the elbow.
• Install the correct backwash flow control.
• Reverse the procedure to reassemble.
15
Note: The clamp and valve will be able to rotate on the tank until pressure is applied.
Note: Make sure to push the valve straight down onto the manifold. If the valve is cocked, it
may cause the o-ring to slip off the manifold.
Figure 6
Note: The number on the flow control should face into the valve body.
Note: Dip switch #5 is to be in the “off” position for 9” and 10” Tanks (0.45 gpm refill
flow control). Dip switch 5 is to be in the “on” position for 12” and 14” tanks (0.80 gpm
brine refill flow control).