Bactrox Operations Manual 18

Problem

Possible Cause

Solution

Amount of Water Collected in the
Dehumidifier:

The amount of water collected in
the dehumidifier depends on the
number and types of samples in
the chamber. The typical amount
collected is around 50 cc each
day of chamber use.

Excessive moisture buildup in
chamber.

A. Thermoelectric condensate

controller fan is not
operating.

A. Replace fan.

B. Ambient room temperature

too low.

B1. Maintain room temperature a

bit higher.

B2. Cover chamber with a blanket

overnight as means of
installation.

C. Air conditioner unit directly

above chamber.

C. Relocate chamber, redirect

the air conditioning duct.

D. Thermoelectric condensate

controller not emptied
regularly.

D. Empty thermoelectric

condensate controller
container.