CAUTION: Extreme caution must be exercised any time access is made into areas housing
electrical components. Repair, replacement or adjustment of components in these areas must
only be done by qualified technicians familiar with electrical circuitry and the operation of the
hypoxic chamber.

ATTENTION: Des Précautions extrèmes sont requises à chaque fois que vous accédez à des
endroits qui abritent les composantes électriques. La réparation, le remplacement oi
ìajustement de composanted dans ces endroits ne doivent être effectués que par un technicien
qualifié familier avec les circuits électriques et ì opération des étuves anaérobiques.

17.

TROUBLESHOOTING

I. PASS BOX

Problem

Possible Cause

Solution

PASS BOX Purge will not start.

A. Input/output terminal buss is not

seated correctly.

A. Press down on the terminal buss

so that it snaps into place.

NOTE: Power should be turned off.

B. User pushed start switch and

held inadvertently held down.
Thus aborting Auto Cycle.

B. Push start switch once then

release.

D. Start button inoperative -

D. Replace start button.

E. No power to controller - check

power light on Controller.

E1. Check for 24 volt DC out of

power supply.

E2. Check for continuity between

power supply and controller.

II. INCUBATOR

Problem

Possible Cause

Solution

Steady digital readout matches
set point, but actual temperature
does not.

A. Temperature Controller out of

calibration.

A. Follow Complete Temperature

Controller instruction.

B. Incubator door open.

B. Close incubator door and

allow incubator to stabilize.

Digital

readout

and

actual

temperature continue to increase
past set point.

A. Output relay stuck - Check to

see if" Incubator Heating"
Indicator light stays on while
temperature is Increasing.

A. Replace Temperature

Controller.

B. Incubator door open.

B. Close incubator door and

allow incubator to stabilize.

Audible alarm sounds.

A. Temperature has risen beyond

the set range for the High
Limit Safety.

A. Determine cause of

temperature increase from
prior section and take
appropriate action.
Call for service.

III. MISCELLANEOUS

Problem

Possible Cause

Solution

Role of Condensate Controller:

The Condensate Controller is an
exclusive Bactrox feature of
Sheldon Manufacturing, Inc. Our
competitors use desiccants.
Desiccants are a drying agent
and often dry out samples.