5 cleaning the electromagnetic valve – LABEC Shinva-45N User Manual
Page 18

16
1 Remove the fastening screw of the safety valve with a screwdriver, take the valve from
base
2 Replace it with a qualified safety valve. (Ensure it opens at a pressure 0.25Mpa-0.26Mpa.)
7.5 Cleaning the Electromagnetic Valve.
1. Dismantle the sterilizer external cover and find broken electromagnetic valve.
2. Poke the pressing disc with screwdriver.
3. Take out the loop of valve.
4. Open the valve with spanner.
5. Clean the valve with water.
6. Restore the valve.
As following :