LABEC Shinva-45N User Manual
Page 10
8
3 The screen will display program parameters and picture as follows after the door is
opened
3.2.3
Vacuum Test
1 Press“▲” key or “▼” key until screen display 6# vacuum test program shows as
following,close the door and press “on/off” key and the vacuum test program will
start.
2 The buzzer alarming will remind user that the vacuum test program is finished, and
screen will display the following:
3
Leakage rate must be lower than 0.13KPa/min。
3.2.4
B&D Test
1 Press“▲”key or“▼” key until screen display 5#B&D test program picture shows as
following:
1#i ns t r ument door : open
St er i l i zat i on t emper at ur e :134. 0℃
St er i l i zat i on t i me: 4. 0 mi nut es
Dr ynes s t i me: 2. 0 mi nut es
6#vacuum
door : open
Test leakage rate
6#vacuum t es t 20:00
Leakage r at e:
0. ***kPa/ Mi n
Pl eas e open t he door
5#BDt es t
door : open
St er i l i zat i on t emper at ur e :134. 0℃
St er i l i zat i on t i me: 3. 5 mi nut es
Dr ynes s t i me: 2. 0 mi nut es