LABEC Shinva-45N User Manual
Page 17

15
Symptom
Possible Cause
Correction
11.2 no printing ink
11.2 change new printing ink
7.2 How to Remove the Cover of Autoclave
!!!
Make sure that the electrical cord is disconnected from the power supply and that there is no
pressure in the chamber.
Removal process:
1
Remove the screws on the side and back
2
Lift up the cover
7.3 How to Check the Safety Valve
In order to prevent the safety valve from blockage, every month under ordinary usage, it is necessary to allow the
steam pressure to escape through it.
1 Run 1# program
2 When pressure of chamber is 210Kpa, pull the ring of safety valve and open it 2 seconds
7.4 How to Replace the Safety Valve
Attention: This method should be used only by professionals.
This manual is related to the following products: