Chapt e r 6 mai nt e nanc e – LABEC Shinva-45N User Manual
Page 14

12
4、 After selecting desired program, press “start” key, when sterilization program is running,
the process are: preparation, pulse vacuum, pulse heating, pulse air exhaust, sterilization,
steam exhaust and pressure release, dryness, balance pressure, sterilization finished.
5、 In autoclave running, the user can observe the parameters information by screen, if the
autoclave is equipped with a printer, all process data can be record by it.
6、 The system will stop running and alarm will appear “surpassed temperature” or “lack of
water”. When these problems appear, please wait for the pressure to balance, buzzer to
sound and screen display “press Esc key”, and then open the door.
7、 When screen display shows “sterilization finished, please open the door” and the buzzer
sounds, the user can open the door and retrieve the articles.
8、 Close the door, turn off the power switch and disconnect power.
5.2 Quit while Autoclave is Running
User can not forcefully open the door when the autoclave is running, if there is a need stop the
program, please press “Esc” and wait for pressure to balance, buzzer to sound, and then press “Esc”
key again to open the door. Screen will go back to the menu picture.
Chapt e r 6 Mai nt e nanc e
The operations described in this chapter must be followed to keep the sterilizer in good
condition and to reduce breakdowns.
!!!
Caution:
Before starting, ensure the electric cord is disconnected and there is no pressure in the
sterilizer.