LABEC Shinva-45N User Manual
Page 13

11
Chapter 5 Operation
5.1 Operation Process
1 Ensure there is enough water in the water tank and that the waste water tank is
empty.
To empty water from waste water tank, use water drain plug in the side of the water
tank and drain water.
If you want to drain water from water tank, the user can lay flat the drain tube.
2 Connect power and switch on, put articles into chamber when screen indicates
ready and then close the door.
3 Select correct program according to articles being sterlizied.
Press “▲” key,screen will in turn display 1#instrument, 2#instrument bag, 3#dressing,
4#rubber, 5#B&D test, 6# vacuum test program and related menu. As following:
Press “set” key, screen display mode shows as following:
Press“▲”key or“▼” key,select pre-heating mode,print mode or date and start to set.
For example, if you want to set the pre-heating mode, the background is clear after this
mode is selected, then press “set” key, the value will flicker and now user can press “▲”
or“▼” to adjust. Press “Esc” when setting is finished.
Program
Sterilization
temperature
Sterilization
time
Dryness time
1# instrument
134
4min
2min
2#instrument bag
134
4min
10min
3# dressing
134
6 min
15min
4# rubber
121
20 min
5min
5# B&D test
134
3.5 min
2 min
6#vacuum test
Test vacuum leakage rate of autoclave
Mode set
Pre-heating mode: Close
Print mode: Open
2008/11/19 09:20