3 improper use – KERN & SOHN MCB 300K100M User Manual
Page 8
When the scale doesn’t connect with the transmission cable,
please do not touch the transmission port to prevent ESD
interference occurs.
The chair scales may not be used for the transport of people!
As long as the patient is remaining on the chair scales, the
wheel brakes must be locked without fail.
Do not step onto the foot rests when stepping or leaving the
chair scales!
3.3 Improper Use
Do not use these scales for dynamic weighing processes.
Do not leave continuous loads on the seating surface or weighing plate. This may
damage the measuring system.
Impacts and overloading exceeding the stated maximum load (max) of the seating
surface or the weighing plate, minus a possibly existing tare load, must be strictly
avoided. This could cause damage to the balance.
Never operate balance in explosive environment. The serial version is not explosion
protected. It should be noted that a flammable mixture of anaesthetics and oxygen or
laughing gas may occur.
Do not modify the construction of the scales. This may lead to incorrect weighing
results, safety-related faults and destruction of the balances.
The balances may only be used according to the described conditions. Other areas
of use must be released by KERN in writing.
8
MCB_MPT-BA-e-1418