CAL U
��
CAL O
��
Press
HOLD
repeatedly until “CAL 0“ appears.
30770
CAL O
��
Press
TARE
,
the display will show a numeric value.
Then press
ENTER
,
“CAL 0“ appears.
CAL 5
��
Press
HOLD
,
“CAL 5“ appears.
200.0
250.0
Press
TARE
Enter the required size of the adjustment weight (See
Section 1, "Technical Data"):
For this select the digit to be changed with
HOLD
and the
numeric value using
TARE
.
Acknowledge by
ENTER
.
82077
Place the adjustment weight carefully in the centre of the
weighing plate; a numerical value will appear in the display.
ENTER
Press
The adjustment process is started.
250.0
The balance will automatically return to Weighing mode upon
successful completion of the adjustment and will display the
weight of the adjustment weight.
Take away adjustment weight.
For verified scales, turn off the scales and move the
adjustment switch back to verification position.
46