Guidance and manufacturer’s declaration-electromagnetic immunity


The MCB 300K100M, MPT 300K100M is intended for use in the electromagnetic environment
specified below.
The customer or the user of the MCB 300K100M, MPT 300K100M should assure that it is used in
such an environment.

Immunity test

IEC 60601 test

level

Compliance level

Electromagnetic

environment-guidance

Electrostatic
discharge(ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%

Electrical fast
transient/burst IEC 61000-
4-4

± 2kV for power
supply lines + 1kV for
input/output lines

± 2kV for power
supply lines Not
applicable

Mains power quality should be
that of a typical commercial or
hospital environment.

Surge IEC 61000-4-5

± 1kV line(s) to line(s)
± 2kV line(s) to earth

± 1kV differential
mode Not applicable

Mains power quality should be
that of a typical commercial or
hospital environment.

Voltage Dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11

<5% UT(>95% dip in
UT) for 0,5 cycle 40%
UT(60% dip in UT) for
5 cycles 70% UT(30%
dip in UT) for 25
cycles <5% UT(>95%
dip in UT) for 5 s

<5% UT(>95% dip in
UT) for 0,5 cycle 40%
UT(60% dip in UT) for
5 cycles 70% UT(30%
dip in UT) for 25
cycles <5% UT(>95%
dip in UT) for 5 s

Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the MCB 300K100M,
MPT 300K100M

requires continued operation
during power mains
interruptions, it is
recommended that the MCB
300K100M, MPT 300K100M

be powered from an
uninterruptible power supply or
a battery.

Power frequency(50/60
Hz) magnetic field IEC
61000-4-8

3 A/m

3 A/m

MCB 300K100M, MPT
300K100M

power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.

NOTE UT is the a.c. mains voltage prior to application of the test level.