30

MBB-M-BA-e-1314

Position of the adjustment switch:

Position of the adjustment
switch

State

1. to left

Not documented

2. concentric

Adjustment position - adjustment possible

3. to right

Verification position - adjustment locked

15.3 Checking the balance verification settings
For the adjustment function, the balance must be switched over to service mode. To

achieve the effect, turn the adjustment switch to adjustment position.

In the service mode all parameters of the balance can be modified. The service

parameters may not be modified, as this could damage the balance settings.

15.3.1 Menu overview in service mode (adjustment switch in adjustment

position)

This overview is merely for checking the parameters set by the appropriate Bureau of

Standards.

Changes may only be made to the parameters for the automatic shut-off function

“ and the audio signal “

“.

1 2 3