15 verification – KERN & SOHN MBB 15K2DM User Manual

Page 27

background image

MBB-M-BA-e-1314

27

15 Verification

Verified scales bear a verification mark or one or more seals affixed by the Bureau of

Standards or the manufacturer on or inside the housing which will self-destroy on

removal. This makes it impossible to verify scales without damaging the seals.

15.1 Adjustment

Observe stable environmental conditions. A warming up time (see chapter 1) is

required for stabilization.

In calibrated balances the adjustment function is switch locked.

In order to carry out adjustments, the switch must be turned to

adjustment position (centre position) (see chapter. 16.2).


Display

Operation

Start balance by pressing

OFF

ON

Press

for approx. 3 sec,

in the display appears „SETUP“, followed by „UNIT“

Press

TARE

repeatedly until “CAL iB“ appears

Press

HOLD

, „CAL U“ appears

Press

TARE

, the appeared triangle  must be located in the

upper left part of the display.

If it is not, press again

TARE

.