14 Instant help

In case of a fault in the program sequence, the balance should be shortly switched

off. The weighing process must then be restarted from the beginning.

Failure:

Possible cause:

The displayed weight

does not glow.

The balance is not switched on.

The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

Check fuse of adapter / glowing green LED next to

fuse

Power supply interrupted.

Batteries are inserted incorrectly or empty

No batteries inserted.

The displayed weight is

permanently changing

Draught/air movement

Table/floor vibrations

The seat surface/weighing plate is in contact with

foreign bodies or is not correctly positioned.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

The weighing result is

obviously incorrect

The display of the balance is not at zero.

Adjustment is no longer correct.

Great fluctuations in temperature.

The balance is on an uneven surface.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

Should other error messages occur, switch balance off and then on again. If the

error message remains inform manufacturer.