KERN & SOHN KDE-T User Manual
Page 17

17
KDE-T-BA_IA-e-1210
7.8
Parts counting
During piece counting parts can either be counted into a container or out of a
container. To count a greater number of parts the average weight per part has to be
determined with a small quantity (reference quantity). The larger the reference
quantity, the higher the counting exactness.
High reference must be selected for small parts or parts with considerably different
sizes.
The average piece weight can only be determined by stable weighing
values.
If weighing values are under zero, the piece counter display shows a
negative number of items.
Reset balance to zero or tare the empty weighing container if necessary.
Press
to show a flashing display of reference number of pieces “5 PCS”
Press
to select further reference numbers of pieces
(example)
Place as many parts to count on the weighing plate as the set reference quantity
requires.
Press
to confirm the selected reference number of pieces and the display
will stop flashing.
The balance is now in parts counting mode and counts all units on the weighing
plate.
The larger the reference quantity, the more accurate the parts
counting.
Pressing
will return the weighing scale to weighing mode and show the weight
for all the pieces counted.
Printout example:
P
2pcs