KERN & SOHN HFC 10T-3 User Manual
Page 30

10. Press
, the display for entering the lower limit value „SETLo“ appears.
11. Confirm by
, the enabled digit flashes.
12. Use
and
to enter lower limit; see step 8.
13. Confirm input by
, “SETLo“ will be displayed.
14. Press
repeatedly to exit menu. From here evaluation takes place
whether the goods to be weighed are within the two tolerance limits.
Start tolerance check: Tare when using a weighing container.
Put on goods to be weighed, tolerance control is started. The signal lights indicate
whether the load is within the two set limits.
Depending on the setting in the menu an audio signal may be sounded.
Load below specified
tolerance
Load within specified
tolerance
Load exceeds specified
tolerance
The indicator [LO] will
appear
The indicator [OK]
appears
The indicator [HI] appears
• The tolerance control is not active when the weight is under 20d.
• To delete limits, enter “0000 kg“.
30
HFC-BA-e-1511