9 peak value function, 10 weighing with tolerance range – KERN & SOHN HFC 10T-3 User Manual
Page 28

5.9
Peak value function
This function displays the highest load value (peak value) of a weighing.
Menu setting of “P4 HLd
PEAK“, see chap. 6
Attention:
Never exceed the maximum permitted load of the peak value on the
balance (!!Danger of breaking!!).
Suspend the material to be weighed
To start the peak load measuring cycle, press
, the heaviest load of the last
weighing activity will be displayed.
The indicator
appears above HOLD.
The peak load will be held on the display until deleted by pressing
.
5.10 Weighing with tolerance range
To ensure that your weight value is within a defined tolerance range, you can define
an upper and a lower limit value of your choice using function “F4 chk“ (See chap. 6).
During tolerance tests such as portioning or sorting the unit will indicate exceeded or
undershot limits by emitting an optical or acoustic signal.
Audio signal:
The audio signal depends on the settings in menu “F0 off
beep“.
Options:
off
Audio signal turned off
ok
The audio signal will be sounded if the load is within the specified tolerance.
ng
The audio signal will be sounded if the load is outside the specified tolerance
28
HFC-BA-e-1511