8.2

Carry out linearization

Turn off balance and attach suitable lifting tackle.

Turn on balance with the lifting tackle attached and press

during self-test. The first function „F0bk“ is displayed.

Press

repeatedly until „F5CAL“ is displayed.

Press

, the last parameter set will be displayed.

• „nonLi“

Adjustment of balance

or

• „LinE“

Linearisation of balance

Select „LinE“ by

Press

, “LoAd 0“ will appear. Make sure that no loads

apart from the lifting tackle are attached to the hook.

Wait for stability display and press

again. “LoAd 1“

(1/3 max) will be displayed.

Attach first adjustment weight, wait for stability display

Press

, “LoAd 2“ (2/3 max) will be displayed.

HFA-BA-e-1511