12 instant help, Instant help – KERN & SOHN VFS 2T1 User Manual
Page 36

VFS-BA-e-1111
36
12 Instant help
In case of an error in the program process, briefly turn off the weighing system and
disconnect from power supply. The weighing process must then be restarted from the
beginning.
Help:
Fault Possible
cause
The displayed weight does not glow.
• Weighing system is not switched on.
• Rechargeable battery empty
The displayed weight is permanently
changing
• Draught/air
movement
-
• Sites with vibration.
• Loading fork is in contact with foreign
matter.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing value is obviously
• The display of the balance is not at zero
wrong
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch weighing system off and then on again.
If the error message remains inform manufacturer.