9.4 Linearization (non-verified models only)
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.
Carrying out linearization is restricted to specialist staff possessing
well acquainted with the workings of weighing scales.
The test weights to be used must be adapted to the weighing scale’s
specifications; see chapter 2.4 “Testing instruments control”.
Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
After successful linearization you will have to carry out calibration;
see chapter 2.4 “Testing instruments control”
Tab. 1: Adjustment points
Adjustment
weight
GAB 6K0.05N
GAB 12K0.1N
GAB 30K0.2N
1.
0 kg
0 kg
0 kg
2.
2 kg
4 kg
10 kg
3.
4 kg
8 kg
20 kg
4.
6 kg
12 kg
30 kg