9.1 Verified models
Switch on device via ON-OFF and press the adjustment
switch at the same time.
After that, during the balance’s selftest, press
TARE
and
%
at the same time. „UnLoAd“ appears.
Acknowledge by
. Ensure that there are no objects on
the weighing plate.
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see operating
instructions chap. 8) select the desired setting, the active
digit is flashing.
Acknowledge by
. „LoAd“ is displayed.
Carefully place adjusting weight in the centre of the
weighing plate. Wait for stability display, then press
.
„PASS“ appears
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, the appliance will
return into weighing mode automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.
GROSS
ZERO