Non-verifiable models – KERN & SOHN GAB 30K5DNM User Manual
Page 18

18
GAB-N-BA-e-1315
9.2 Non-verifiable models:
Switch on balance Whilst balance is carrying out self-test
(counts down from 99… to 00…)
TARE
and
%
press at the
same time.
“UnLoad“ will be displayed, follo
wed by flashing numeric
value.
(Example)
Change by pressing the navigation keys (See manual chap.
8); select desired setting; currently enabled digit will be
flashing.
(Example)
Confirm value by
0
, “Load“
will be displayed.
Place adjustment weight, stability display appears
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during self test,
balance will return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting weight
the display will show an error message, repeat adjustment
process.
GROSS
ZERO