beautypg.com

Using the jb aqua plus, Bath controls, Indicator lamps – Grant Instruments SBB Aqua Plus User Manual

Page 8: Setting the control temperature)

background image

29825 Ver. 2 - July 2010 71W

SUB, JB & SBB Aqua Plus

Page 7

Operating Manual

www.grant.co.uk


5.2

Using the JB Aqua Plus


5.2.1 Bath controls


5.2.2 Indicator lamps

There are three indicator lights:
1. Power on (green).
2. Heater on (orange). Marked

.

3. Warning (orange). Marked indicates that the temperature is being controlled by the

sample protection thermostat.


5.2.3 Setting the control temperature ()

The water temperature of the bath can be set using the larger temperature control knob.

1. Turn the knob of the sample protection thermostat fully to maximum.
2. Turn the knob of the temperature control to the desired temperature.
3. Turn on the unit using the switch on the rear of the bath and wait until the

temperature has stabilised.

4. Measure the temperature with a thermometer and adjust the temperature control

knob if necessary to obtain the desired temperature.

5. Set the sample protection thermostat as detailed below if needed.

Sample

protection

thermostat

Power on light

Heater on light

Temperature control

Warning light