beautypg.com

Resetting the over temperature cut-out devices, 0 technical specifications – Grant Instruments SBB Aqua Plus User Manual

Page 15

background image

SUB, JB & SBB Aqua Plus

29825 Ver. 2 - July 2010 71W

Operating Manual

Page 14

www.grant.co.uk

EN


After filling the bath adjust the water flow rate to the minimum which maintains a constant
water level when the water is boiling. Check that any relevant local water supply
regulations are complied with if connecting to a mains water supply.

5.4.5 Resetting the over temperature cut-out devices

Two fixed over temperature cut-outs prevent the heater from overheating in the case of a
low water level. The reset buttons for the over temperature cut-outs are on the side panel
and can be reset by removing the protective black cap and pressing the red button in.

The over temperature cut-outs may activate during normal operation. If this is the case
then they can be reset and the bath can continue to be used without compromising the
safety of persons or the surroundings. However, if the cut-outs are persistently activated
then it is recommended to have the bath checked by a competent person as soon as
possible.

6.0 Technical specifications


Operating conditions

Ambient Temperature

5 to 40°C

Maximum relative humidity

80% R.H. in room temperatures up to 31°C
decreasing linearly to 50 % R.H. at 40°C

Altitude above sea level

Up to 2,000 m (6,500 ft)

Operating Enviroment

Indoor use only



Electrical details

Mains supply:

230V @ 50/60 Hz or 120V @ 50/60 Hz

Polution degree:

2

Installation Category: II

Note: Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply
voltage

Models

Capacity (L)

Current Rating (A)

120V

230V

SUB Aqua 2 Plus

2

1.0

0.5

SUB Aqua 2S Plus

2

2.6

1.3

SUB Aqua 5 Plus

5

2.6

1.3

SUB Aqua 12 Plus

12

5.3

2.7

SUB Aqua 18 Plus

18

7.9

5.3

SUB Aqua 26 Plus

26

7.9

5.3

SUB Aqua Dual Plus

5 & 12

7.9

4.0

SUB Aqua 34 Plus

34

10.5

8.5

JB Aqua 2 Plus

2

1.0

0.5

JB Aqua 2S Plus

2

2.8

1.3

JB Aqua 5 Plus

5

2.8

1.3

JB Aqua 12 Plus

12

5.6

2.7

JB Aqua 18 Plus

18

8.4

5.3

JB Aqua 26 Plus

26

8.4

5.3

JB Aqua Dual Plus

5 & 12

8.2

4.0