beautypg.com

Flintec – Flintec FAD-4 User Manual

Page 16

background image

FLINTEC

Document

Date

Code

Rev.

Page

FAD - 4 DIGITAL LOADCELL JUNCTION BOX

02.05.00

FAD4_2RM

2

16 of 20

5.2 ELECTRONIC

(mV/V).

Assume a scale of Max= 30/60kg , e=0.010/0.020kg , with 4 loadcells of rated capacity 50kg -2mV/V each
and the dead load is 1.940kg.
Conversion rate is 5/sec , the digital filter is normal .
Loadcell data:
L/C1 : 1.9793 mV/V output at 50kg and zero balance : 0.0257 mV/V.
L/C2 : 1.9392 mV/V output at 50kg and zero balance : 0.0276 mV/V.
L/C3 : 1.9577 mV/V output at 50kg and zero balance : 0.0553 mV/V.
L/C4 : 1.9640 mV/V output at 50kg and zero balance : -0.0022 mV/V.

SPAN average = (1.9793 + 1.9392 + 1.9577 +1.9640) / 4=1.9600 mV/V

L/C ZERO BALANCE average = [0.0257 + 0.0276 + 0.0553 + (- 0.0022)] / 4 = 0.0266 mV/V

SCALE DEDLOAD : 1.9600 mV/V . 1.940 kg = 0.0190 mV/V

(4x50kg)

OVERALL DEADLOAD = L/C ZERO BALANCE + SCALE DEADLOAD = 0.0456 mV/V

The calibration procedure should be as follows :
-

Power on the FAD-4 .

Send CORNER DEFINITION commands:

STX ADD C 1 019793 BCS2 BCS1 ETX CORNER

1

STX ADD C 2 019392 BCS2 BCS1 ETX CORNER

2

STX ADD C 3 019577 BCS2 BCS1 ETX CORNER

3

STX ADD C 4 019640 BCS2 BCS1 ETX CORNER

4



Send DEAD LOAD calibration command :

STX ADD A 000456 BCS2 BCS1 ETX



Send SPAN calibration command :

STX ADD B 50000 BCS2 BCS1 ETX


RATED

CAPACITY

OF

EACH

LOADCELL



Send STORE command to save in EEPROM

STX ADD W BCS2 BCS1 ETX

- End.