8.10. Save Calibration and Setup Data Commands – CS, WP
The calibration and setup parameters can be divided in 2 groups:
Calibration
: CMn, CI, MR, DS, DP, CZ, CG, ZT, ZR, FD, LC and LNn, saved by command CS
Setup
: FL, FM, NR, NT, BR, AD, DX and other, saved by command WP
Note:
Calibration data can only be saved if the TAC code is known and precedes the CS command. See the
CE
and CS commands in chapter 8.2.
The setup data are stored non-volatile in the EEPROM using the WP command.
8.10.1. WP Einstell-Parameter sichern
With this command the settings of the “Filter” (FL, FM) ,the “No-motion” (NR, NT) and the communication (AD,
BR, DX) will saved in the EEPROM.
Master (PC / SPS) sends
Slave (LDU XX.X) responds
Meaning
WP
OK
Save setup data
WP
ERR
Error