8.10. Save Calibration and Setup Data Commands – CS, WP

The calibration and setup parameters can be divided in 2 groups:

Calibration

: CMn, CI, MR, DS, DP, CZ, CG, ZT, ZR, FD, LC and LNn, saved by command CS

Setup

: FL, FM, NR, NT, BR, AD, DX and other, saved by command WP

Note:

Calibration data can only be saved if the TAC code is known and precedes the CS command. See the

CE

and CS commands in chapter 8.2.

The setup data are stored non-volatile in the EEPROM using the WP command.

8.10.1. WP Einstell-Parameter sichern

With this command the settings of the “Filter” (FL, FM) ,the “No-motion” (NR, NT) and the communication (AD,
BR, DX) will saved in the EEPROM.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

WP

OK

Save setup data

WP

ERR

Error