8.2.13. LC

Clear Linearisation

This command puts the LDU back to the initial factory calibration using two calibration points, zero at 0.0 mV/V
and 200 000 d at 2.0 mV/V.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

LC

OK Linearize clear

8.2.14. LN n Linearisation

This command makes it possible to inspect and correct more than the two normal nodes set by the CZ and CG
command – up to seven nodes are allowed. Issuing the command with only one parameter (1-7) returns the
node in the format: Ln:XXXXXXX YYYYYYY where XXXXXXX is the input level in mV/V (without decimal point)
and YYYYYYY is the displayed number of increments.
The calibration can be done in the following way: LN1 0 0 defines the zero output at zero mV/V input. LN2
200000 100000 defines 100 000 d at +2mV/V input – at least two nodes must be given.

The following table shows a linearization with 3 nodes

.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

LN 1 0 0

OK

Setup: Node 1 = 0 mV/V = 0 d

CE 17

OK

Calibration sequence active

LN 2 100000 100000

OK

Setup: Node 2 = 1 mV/V = 100 000 d

CE 17

OK

Calibration sequence active

LN 3 200000 200000

OK

Setup: Node 3 = 2 mV/V = 200 000 d

CE 17

OK

Calibration sequence active

CS

OK

Calibration values saved

LN 3

L3:+200000+200000

Request: Node 3 = 2 mV/V = 200 000 d

Note:

Before you re-calibrate with less calibration points use the command LC (as described before).

Factory default: LN = 2 calibration points with 0 mV/V = 0 d and 2 mV/V = 200 000 d.

Hint:

By using the load cell’s characteristic data the linearization allows for an electronic calibration (eCal).

8.2.15. CS

Save the Calibration Data

This command results in the calibration data being saved to the EEPROM and causes the TAC to be
incremented by 1.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

CS

OK

Calibration values saved

The CS command saves all of the calibration group values, as set by CZ, CG, CM n, DS, DP and ZT. The
command returns ERR and has no updating action unless it is preceded by the CE_XXXXX.