beautypg.com

Commands – overview – Flintec LDU 69.1 User Manual

Page 12

background image

LDU 69.1 Technisches Handbuch, Rev. 12 January 2011

Page 12 of 28

7.

Commands – Overview

Command Short

description

Parameter

value

Page

AD

Communication: Device Address

0...255

24

BR

Communication: Baud Rate

9600…115200 baud

24

CE

Calibration: Open Calibration Sequence; Read TAC Counter

0...65535

14

CG

Calibration: Set Calibration Gain (span) at load > zero

0...999999

16

CI

Calibration: Minimum Output Value

–999999...0 d

14

CL

Communication: Close Device

None

24

CM n

Calibration: Set Maximum Output Value (n = 1, 2 or 3)

1...999999 d

14

CS

Save the calibration data (CMn, CI, DS, DP, CZ, CG, u.a.) to EEPROM None

17, 25

CZ

Calibration: Set Calibration Zero Point – scale without load

None

15

DP

Calibration: Set Decimal Point Position

0...5

15

DS

Calibration: Set Display Step Size

1, 2, 5, 10, 20, 50,
100, 200

15

DX

Communication: Set Half-duplex (0) or Full-duplex (1)

0 or 1

24

FD

Factory default settings: Write data to EEPROM (TAC protected)

None

16

FL

Digital filter: Filter Settings

0...17

19

GG

Output: Get Gross Value

None

21

GH

Output: Get Hold Weight

None

22

GN

Output: Get Net Value

None

21

GS

Output: Get ADC Sample Value

None

21

GT

Output: Get Tare Value

None

21

GW

Output: Get Data String “Net/Gross/Status“

None

21

HW

Output: Hold weight for time-synchronous capturing

None

22

ID

Device information: Identify Device

None

13

IS

Device information: Identify Device Status

None

13

IV

Device information: Identify Firmware Version

None

13

LC

Linearisation: Clear data

None

16

LN

Linearisation: Define up to 7 calibration points

0...200000 d

17

MR

Calibration: Define Multi-interval (0) or Multi-range (1)

0 or 1

15

NR

Motion detection: No-motion Range

0...65535 d

18

NT

Motion detection: No-motion Time Period

0...65535 ms

18

OP

Communication: Open Device xxx

0...255

24

RS

Device information: Read serial number

None

13

RT

Scale function: Reset tare and switch to gross indication

None

20

RZ

Scale function: Reset zero point

None

20

SG

Auto-transmit: Send Gross Value continuously

None

23

SN

Auto-transmit: Send Net Value continuously

None

23

SR

Reset firmware (warm start)

None

13

ST

Scale function: Set Tare and switch to Net Indication

None

20

SW

Auto-transmit: Send Data String „Net/Gross/Status“ continuously

None

23

SZ

Scale function: Set Zero

None

20

UR

Digital filter: Update Rate

0...7 (= 2

0

bis 2

7

)

19

WP

Save the Setup Data (FL, NR, NT, AD, BR, DX) to EEPROM

None

25

ZR

Calibration: Set Zero range

0...999999

16

ZI

Calibration: Initial Zero Range

0...999999 d

16

ZT

Calibration: Disable (0) or enable (1) Zero tracking

0 or 1

16