Synchronized output commands – hw, gh – Flintec LDU 69.1 User Manual
Page 22
LDU 69.1 Technisches Handbuch, Rev. 12 January 2011
Page 22 of 28
8.7. Synchronized Output Commands – HW, GH
For systems requiring simultaneous sampling of data two commands are available:
8.7.1. HW Hold
Weight
Issuing the HW command, which has no parameters, will latch the current net weight in a register for later
readout in all LDU’s in a common network regardless of their individual address.
Note:
The LDU’s will not respond to the HW command on the interface lines – it would lead to bus contention
Master (PC / SPS) sends
Slave (LDU XX.X) responds
Meaning
HW
No
response
Hold weight value in register
8.7.2. GH Get
Hold
Weight
Issuing the GH command, which has no parameters, will return the current hold weight in the format N+000100
from the currently addressed LDU.
Master (PC / SPS) sends
Slave (LDU XX.X) responds
Meaning
OP 1
OK
Open device #1
GH
H+01.100
Hold weight value: 1,100 d
All other channels (LDU with other address) have to be requested one by one.