LDU 69.1 Technisches Handbuch, Rev. 12 January 2011

Page 16 of 28

8.2.8. CG Set Calibration Gain (Span)

This is the reference point for calibration under load, and is subject to TAC control.

Permitted values are from 1 to 999 999.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CG

G+10000

Request: Calibration weight = 10 000 d

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

CG 15000

OK

Setup: Calibration weight = 15 000 d

For calibration an input signal near the display maximum (CM n) will give the best system performance. The
minimum calibration load of at least 20% is recommended. Is the calibration weight smaller than 1% of display
maximum (CM n), the LDU will respond with an error message (“ERR”).
Factory default: 20 000 = 2.000 mV/V input signal

8.2.9. ZT

Enable / Disable Zero Tracking

This command enables or disables the zero tracking. ZT = 0 disables the zero tracking and ZT = 1 enables the
zero tracking. Issuing the command without any parameter returns the current ZT value.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

ZT

Z:001

Request: ZT status

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

ZT 0

OK

Setup: ZT = Disable

Zero tracking will be performed only on results less than +/-0.5 d at a rate of 0.4 d/sec, where d = display step
size (see DS command). The zero can only be tracked to +/- 2% of maximum (see CM n command).
Factory default: ZT = 0 [Disable]

8.2.10. ZR

Set Zero Range

This command defines the zero range. Permitted values are from 0 to 999 999.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

ZR

R+000100

(example)

Request: ZR = 100 d

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

ZR 1000

OK

Setup: ZR = 1 000 d

8.2.11. ZI

Initial Zero Range

Defines the initial zero range (0…999999 d). If ZI is non-zero the device will perform an automatic Set-Zero
when the weight stabilizes with the No-motion settings and the weight is within the ZI range. Factory default: 0.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

ZI 100

OK

Setup: Initial Zero range = 100 d

8.2.12. FD

Reset to Factory Default Settings

This command puts the LDU back to a known state. The data will be written to the EEPROM and the TAC will
be incremented by 1.

Note:

All calibration and setup information will be lost by issuing this command!

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

FD 0

OK

Factory default setting