ALTANA Z7-Z10 Bubble Viscometers User Manual
Page 3

DIRECTIONS FOR USING THE 
GARDNER BUBBLE VISCOMETER A-5 to A-1 
CAT. NO. PV-0500
This instrument is based on the principle that the 
viscosity is directly proportional to the bubble 
speed. The viscosity of a sample of varnish or 
other transparent liquid is therefore determined 
by reference to the standard tubes in which an air 
bubble rises with the same speed as it does in a 
tube of the sample being tested. For this purpose 
two carefully calibrated empty tubes are furnished 
with each set.
The bubble size of the sample is first adjusted as 
nearly as possible to that in the standard tubes. 
The tubes are then brought to a temperature of 
25ºC (77ºF). This is essential due to the difference 
in the temperature coefficient of viscosity of 
mineral oils and varnishes. It is suggested that 
a water bath be used, which quickly brings the 
tubes to the proper temperature and insures 
greater accuracy.
The bubble should be started from the flat end 
of the tube held in a true vertical position. The 
bottom meniscus of the bubble should be used 
for comparing the samples. It is only necessary 
that they start together. Owing to unavoidable 
differences in bubble sizes and in the shapes of 
the ends of the tubes, it may be necessary to raise 
one of the tubes so that the points of comparison 
will be on the same level. The observation is made 
after the bubbles are completely formes and 
moving upwards along a true vertical line. This 
takes place about 1½ to 2 cm from the bottom of 
the tube.
The standard empty grade “A” tubes have been 
carefully selected to and I.D. of 10.65 ± 0.0025 
mm. and should be used for exact determinations. 
However, this close tolerance is relatively 
expensive, and for the sake of economy, tubes 
Directions for Using the Gardner Bubble Viscometer
2
