2.0 Test Procedure

The indentation hardness test
shall be carried out at an ambient
temperature of +23 °C ± 2 °C and
relative humidity of ( 50 ± 5) %.

Different paint systems need different
drying time, allow enough time to dry,
at leat 16 hours, then check as quick
as possible.

• Put the test panel on a firm,

horizontal surface. Place the
indentation tester carefully and
without any tilting or lateral
movement onto the paint film.
Set down first the feet and then
the indentor for ( 30 ± 1) s.


• The location of the indentation

mark is much easier when the
position of the feet is marked
before the test is started
without touching the tester (see
figure 2).

• Increase the indentation tester

from the specimen in the opposite
sequence, i.e. indentor first,
feet last.

1 Position of Indentation
2 Position of the feet

2

2

1