ALTANA Hoffman Scratch Hardness Tester Manual User Manual
Page 7

6
Operation
Before initial setup, please remove protective film 
from the scale arm (balanced beam). In operation, 
small and/or large riders are attached to this arm 
at the numbered positions. These must be placed 
on the scale very carefully so that the test surface 
is not marred before actual testing is begun. The 
large rider generates a load of 100 grams per 
division, while the small rider generates a load of 
25 grams per division. The small quarter rider may 
be used for marking low range measurements 
involving small increments of pressure, or it may 
serve as a vernier in conjunction with the large 
rider in making more precise medium range 
measurements. Two large riders may be used 
for making high range measurements. The sum 
of all the rider settings represents the total load 
in grams applied by the lever system to the 
scratching tool. The scale arm is counterpoised 
and its graduation marks can be used with any 
combination of riders to create a reproducible load 
on the pivoted lever system. In order to assist you 
in selecting the correct rider for your application, 
listed below are several examples:
• Small quarter rider at position 
 
4: 4 x 25 = 100 grams
• Large standard rider at position 
 
20: 20 x 100 = 2000 grams
• Large standard rider at position 6, plus small ...
 
quarter rider at 2: (6 x 100) + (2 x 25) =
600 + 50 = 650 grams
• Large standard rider at position 20, plus ..........
 
.............another large standard rider at position
10: (20 x 100) + (10 x 100) =
2000 + 1000 = 3000 grams
After attaching the rider(s), place the Hoffman 
Scratch Tester on a flat test surface that measures 
at least 2 x 4 inches (5.1 x 10.2 cm). Place your 
Operation