Technical data – ALTANA BYK-mac COLOR Manual User Manual
Page 52

51
14. Technical Data
General technical data:
Temperature range
10 °C to 42 °C (50 °F to 110 °F)
for operation
0 °C to +60 °C (32 °F to 140 °F )
for storage
Rel. humidity
Up to 85% non-condensing/35°C (95 °F)
Measurement unit:
Color
Measuring Geometry
Illumination 45°,
-15°,15°, 25°, 45°, 75°, 110° aspecular viewing
Measuring Area
Ø23mm (0.91 in.) or 12mm (0.47 in.)
Spectral Range
400 - 700 nm, 10 nm resolution
Measurement Range
0 to 400 % reflectance
Measuring time
BYK-mac
< 6 seconds
BYK-mac COLOR
< 4 seconds
Repeatability
0.02 dE*, 1
10 consecutive readings on white tile
Reproducibility
0.2 dE*, 1
average on 12 BCRA II tiles
Color scales
dE*, dE CMC, dE 94,dE DIN 6175
Illuminants
A; C; D50; D65; F2; F7; F11; F12
Observer
2°; 10°
Effect
only with BYK-mac
Measurement Geometry 15°, 45°, 75° and diffused illumination
perpendicular viewing
Effect parameters
dS, dS_a, dS_i, dG
Repeatability
S_a / S_i: 5% or > 0.50 / G = ± 0.05, 1
Reproducibility
S_a / S_i: 10% or > 1.00 / G = ±0.15, 1
Memory
1000 Standards/Samples
Power supply
Rechargeable battery pack or 4 mignon AA batteries
(alkaline or rechargeable)
Dimensions (LxWxH)
218 x 81 x 147mm (8.6 x 3.2 x 5.8 in.)
Weight
approx. 1300g/ 2.86 lbs
Interface
USB-B
Technical Data