5. OPERATION

5.1. Basic weighing operation

1. Touch the

switch to turn the power ON.

All the symbols except the comparator LEDs are
displayed. When the weight values becomes
stable, the display turns off for a moment and
displays zero (power-on zero) with the weighing
unit used last before turning off.

2. Select a weighing unit using the

UNITS

switch.

See “6. SELECTING A WEIGHING UNIT” in

detail.

3. When the display doesn’t show zero, tap the

switch to set the display to zero.

4. When using a tare (container), place the

container on the weighing pan, and tap the

TARE

switch to set the display to zero.

5. Place the object to be weighed on the weighing

pan or in the container, and wait for the STABLE
indicator to turn on and read the value.

6. Remove the object from the weighing pan.

7. Touch and hold the

switch. Then, the

scale displays “

0ff

” and turns the power off.



5.2. Notes about operations

Power-on zero

If the weight is unstable at power ON, the scale displays “------”. Check anything

touches the weighing pan, or check if there is strong wind or vibration.

The range for power-on zero is within ±50% (±10% for Legal for Trade models) of

the weighing capacity (kg) at the calibrated zero point. If the scale is powered on
with a load beyond this range, the scale displays “------”. Remove the load from
the weighing pan.

ZERO and TARE

The

(as ZERO switch) and

TARE

switches work when the weight value

is stable.

The

switch will zero the scale if the weight value is within ±2% of the

weighing capacity (kg) at the power-on zero point. The ZERO◄ indicator turns on.
(ZERO operation)