10.4. Restoring the calibration data to the factory set values

If the gravity acceleration value or calibration data is changed unintentionally, restore
those values to the factory set values, as follows.



1. Refer to “10.1. Calibration mode” to enter the

calibration mode. The gravity acceleration value is
displayed.




2. Tap the

UNITS

switch twice to display “

ClrC

”.



3. Tap the

TARE

switch to display

ClrC no

” with

no

” blinking.



4. Tap the

SAMPLE

switch.

ClrC no

” changes to “

ClrC go

” with “

go

blinking.

To cancel the restoring procedure, tap the

switch. The display returns to step 2.

5. When “

ClrC go

” is displayed, tap the

TARE

switch. The factory set values are restored

and “

end

” is displayed.

Then, the display returns to “

ClrC

To finish the setting procedure, touch the

switch or press the

CAL

switch. The scale turns

the power OFF.






Tap

Tap

Tap

Tap twice