10.4. Restoring the calibration data to the factory set values
If the gravity acceleration value or calibration data is changed unintentionally, restore
those values to the factory set values, as follows.
1. Refer to “10.1. Calibration mode” to enter the
calibration mode. The gravity acceleration value is
displayed.
2. Tap the
UNITS
switch twice to display “
ClrC
”.
3. Tap the
TARE
switch to display
“
ClrC no
” with
“
no
” blinking.
4. Tap the
SAMPLE
switch.
“
ClrC no
” changes to “
ClrC go
” with “
go
”
blinking.
To cancel the restoring procedure, tap the
switch. The display returns to step 2.
5. When “
ClrC go
” is displayed, tap the
TARE
switch. The factory set values are restored
and “
end
” is displayed.
Then, the display returns to “
ClrC
”
To finish the setting procedure, touch the
switch or press the
CAL
switch. The scale turns
the power OFF.
Tap
Tap
Tap
Tap twice