Troubleshooting, Physical checks – Hardy Advantage HI LPRC03 Load Point User Manual
Page 15
![background image](/manuals/657651/15/background.png)
OPERATION AND INSTALLATION MANUAL
Page 9
NOTE:
For additional help with the load point installation,
contact your local Hardy Instruments Representative
or either Application Engineering or Customer Sup-
port at the Hardy Instruments company.
Troubleshooting
Physical Checks
Follow these steps before runing any electrical tests:
Step 1.
Visually inspect each load point assembly
for physical damage. If you find any prob-
lems, replace the part that is damaged.
•
Look for distortions or cracks in all
metal parts. Make sure that any welds
are not cracked and do not have deep
pot marks.
•
Check all cables for cracks, cuts or
crimping. Check for any abrasions on
the cables.
•
Look for structural changes in the scale
or supporting structures.
•
Look for binding of any kind on the load
point assembly.
Step 2.
In the Troubleshooting section of the Hardy
Manual for the weighing instrument, look
for information on how to troubleshoot
using Integrated Technician. (See also the
support page of the Hardy Web site at: http:/
/www.hardyinst.com.)
Step 3.
Get the Load Sensor certification sheets for
referencing while troubleshooting. They are
also available 24 hours a day at our Web
Site: http://www.hardyinst.com.