beautypg.com

Hardy HI 1756-FC User Manual

Page 39

background image


A1 - Guidelines for Instabilities on Formerly Operating System

UNSTABLE RATE CONTROLLER

Disconnect external signal cables and

shields, except AC Power

Monitor the reading

for stability

STABLE?

Reconnect signal

cables one at a

time

A1

If installing any cable causes

unstable readings

REVIEW TROUBLESHOOTING

SECTION B: B1.1-B1.7

Go to

B

Yes

No

Monitor the

display for stability

Problem could be in

the instrument. Contact

Hardy Customer Support

PH: 800-821-5831

34