beautypg.com

Hardy HI 1756-FC User Manual

Page 19

background image


a 5-point rate calibration. The other points are filled
with data based on these 2 points.

RATE_CAL_5POINT_CMD

0x60

Runs the 5-point rate calibration at whatever points
have previously been defined.

RATE_CAL_2POINT_CMD

0x61

Runs the 2-point rate calibration at whatever points
have previously been defined.

CALLOWCMD

0x64

Sets the “calLowCount” parameter to the current
A/D average counts when doing a hard calibration.
Calls an Integrated Technician function during low
calibration.
Error Return Values:
STATUSWORD - There was a conversion error or
A/D conversion failure.
HARDCALFAILCOUNTS - There are less than
1000 counts between the zero and the span weights.

CALHIGHCMD

0x65

Sets the Span Weight parameter to the current A/D
average counts when doing Hard CAL.
Error Return Values:
STATUSWORD - There was a conversion error or
A/D conversion failure.
HARDCALFAILCOUNTS - There are less than
1000 counts between the zero and the span weights.

C2CALCMD

0x66

Performs a C2 Calibration. Uses CalLowWeight as
the reference point.

Error Return Values:
STATUSWORD - There was a conversion error or
A/D conversion failure.
C2FAILNODEVS – Detected no C2 load cells.
C2FAILCAPEQ - Detected two load cells with
different capacities.

WRITEPARAM

0x67

Writes the parameters. See Parameter Table
below.
Return Data
:
COMMAND STATUS - The status is 0 if the
command succeeded. If a range error was detected
in one of the parameters, a non-zero value is
returned indicating which parameter failed its range.
See Status Table below.

READPARAM

0x69

Reads the parameters. See Parameter Table
below.
Return Data:
COMMAND STATUS See Status Table below.
PARAMETERS

SETDEFAULTPARAMS

0x94

Loads the default settings into RAM, and non-
volatile RAM. Error Return Values: None

14

Chapter 4