Hardy HI 1756-FC User Manual
Page 19
a 5-point rate calibration. The other points are filled
with data based on these 2 points.
RATE_CAL_5POINT_CMD
0x60
Runs the 5-point rate calibration at whatever points
have previously been defined.
RATE_CAL_2POINT_CMD
0x61
Runs the 2-point rate calibration at whatever points
have previously been defined.
CALLOWCMD
0x64
Sets the “calLowCount” parameter to the current
A/D average counts when doing a hard calibration.
Calls an Integrated Technician function during low
calibration.
Error Return Values:
STATUSWORD - There was a conversion error or
A/D conversion failure.
HARDCALFAILCOUNTS - There are less than
1000 counts between the zero and the span weights.
CALHIGHCMD
0x65
Sets the Span Weight parameter to the current A/D
average counts when doing Hard CAL.
Error Return Values:
STATUSWORD - There was a conversion error or
A/D conversion failure.
HARDCALFAILCOUNTS - There are less than
1000 counts between the zero and the span weights.
C2CALCMD
0x66
Performs a C2 Calibration. Uses CalLowWeight as
the reference point.
Error Return Values:
STATUSWORD - There was a conversion error or
A/D conversion failure.
C2FAILNODEVS – Detected no C2 load cells.
C2FAILCAPEQ - Detected two load cells with
different capacities.
WRITEPARAM
0x67
Writes the parameters. See Parameter Table
below.
Return Data:
COMMAND STATUS - The status is 0 if the
command succeeded. If a range error was detected
in one of the parameters, a non-zero value is
returned indicating which parameter failed its range.
See Status Table below.
READPARAM
0x69
Reads the parameters. See Parameter Table
below.
Return Data:
COMMAND STATUS See Status Table below.
PARAMETERS
SETDEFAULTPARAMS
0x94
Loads the default settings into RAM, and non-
volatile RAM. Error Return Values: None
14
●
●