Load cell excitation, C2 calibration output, Number of channels – Hardy HI 1756-FC User Manual
Page 10: Update rate, Averages, Waversaver, Digital voltmeter (it), Power and utility requirements, Voltage, Frequency

1000’ for C2, Non C2, or JB Card
250’ for IT
Load Cell Excitation
5 VDC +/- 1.15 W maximum.
Isolation from digital section 1000 VDC minimum
C2 Calibration Output
Isolation from digital section 1000 VDC minimum
Number of Channels
1 Channel
Update Rate
100 Updates per Second (10 ms)
Averages
1-255 User-selectable in Single Increments
WAVERSAVER®
User Selectable
OFF
7.50 Hz
3.50 Hz
1.00 Hz (Default)
0.50 Hz
0.25 Hz
Digital Voltmeter (IT)
Accuracy ± 2% of full scale
Power and Utility Requirements
Voltage
24 VDC (Standard)
100/240 VAC Universal (Optional)
Frequency
47/63 Hz
Power
10 Watts maximum with options
Environmental Requirements
Digital Voltmeter
Accuracy +/- 2% of full scale
Operating Temperature Range
0 to 60º C (32º F to 140º F)
5
●
●
Chapter 2
Temperature Coefficient
●
●
●