8 cleaning, 1 grinding elements, 2 mill – FRITSCH PULVERISETTE 4 classic line User Manual
Page 45: Cleaning 8.1 grinding elements 8.2 mill, 8cleaning
8
Cleaning
DANGER!
Mains voltage!
– Before beginning with cleaning work, disconnect the
mains plug and protect the device against being unin-
tentionally switched back on!
– Do not allow any liquids to flow into the device.
– Indicate cleaning work with warning signs.
– Put safety equipment back into operation after
cleaning work.
When cleaning the entire device, adhere to the guidelines of
the Accident Prevention Regulation (BGV A3) ‐ especially
when the device has been set up in a dusty environment or
when the grinding stock processed produces dust.
8.1 Grinding elements
NOTICE!
Cool grinding elements made of agate, sintered corundum,
zirconium oxide and silicon nitride slowly and carefully.
Do not heat agate elements in a microwave under any cir-
cumstances (heating is too fast).
They must never be exposed to thermal shocks as this
could cause irreparable damage to the parts ® They will
burst apart like in an explosion.
n
Clean the grinding bowl and grinding balls each time after using
them: Clean them, e.g., under running water using a brush and a
commercially available cleaning agent.
n
Roughly half fill the grinding bowl with grinding balls and a little sand
and water and run for 2 to 3 minutes (correctly tensioned) in the
Vario planetary mill.
n
Cleaning with an ultrasonic cleaner is permitted.
n
For sterilisation in the heat cabinet, only heat the grinding elements
up to 100 °C.
8.2 Mill
n
The Vario planetary mill can be wiped down with a damp cloth after
it has been switched off.
Cleaning
- 45 -