beautypg.com

2 gassing, Accessories – FRITSCH PULVERISETTE 4 classic line User Manual

Page 42

background image

Using the additional clamping system, the following steps can also be

completed in the glove box with subsequent locking into the Vario-

Planetary Mill:

n

Connect the gassing hose (a) to the inert gas supply using the pro-

vided hose clamp (b).

n

Screw the ventilation attachment (e) onto one of the two valves (d).

n

Place the coupling (c) of the gassing hose on the free valve. When

doing so, press the lever of the coupling and push the coupling along

the neck of the valve until it goes no further. Release the lever.

Using the additional clamping system, the closed grinding

bowl can also be tensioned and gassed outside of the

device. (See Ä Chapter 7.1.1 ‘Locking the additional

clamping system into place’ on page 39)

7.1.2.2

Gassing

n

Slowly open the inert gas supply.

n

Press a thin object (e.g. hex key) onto the top of the ventilation

attachment (e) so that the air can escape from the grinding bowl.

n

The inert gas now purges the air from the grinding bowl.

n

The duration of purging has to be determined through experimenta-

tion.
It depends on grinding bowl size, filling, and gas supply, among other

factors.

n

To end purging, close the inert gas supply and release the ventilation

attachment.

n

Screw off the ventilation attachment.

n

Remove the gassing hose. To do so, press the lever.

CAUTION!
Only switch on the device when both coupling and ventila-

tion attachment have been removed.
Overpressure may occur during grinding!

Accessories

- 42 -