Checkline PM4 User Manual
Page 4
Electromatic Equipment Co., Inc
-4-
4.0
CONTACT-FREE TEMPERATURE SENSOR
The back side of the humimeter PM4 contains
an embedded temperature sensor for the
contact-free temperature measurement below
the sensor. This ensures the determination of
the temperature of the product within seconds
and
consequently
the
temperature
compensation of the displayed water content.
The sensor must not be covered by a hand or
any other object!
5.0
CALIBRATION CURVES
name
paper type
density
600 paper
Very low density paper
600 kg/m³
650 paper
Low density paper
650 kg/m³
700 paper
Low density paper
700 kg/m³
750 paper
Fluting, Schrenz
750 kg/m³
800 paper
Newsprint paper
800 kg/m³
850 paper
Kraftliner brown
850 kg/m³
900 paper
Kraftliner White Top, Testliner brown
900 kg/m³
950 paper
Testliner white, copy paper
950 kg/m³
1000 paper
Copy paper, LWC raw
1000 kg/m³
1050 paper
Copy paper calendered
1050 kg/m³
1100 paper
Copy paper calendered
1100 kg/m³
1200 paper
LWC calendered
1200 kg/m³
1300 paper
Flysheet paper
1300 kg/m³
1400 paper
Flysheet paper
1400 kg/m³
1500 paper
Flysheet paper high density
1500 kg/m³
1600 paper
Flysheet paper very high density
1600 kg/m³
Empty 1700
For creation of customer calibration curve
Empty 975
For creation of customer calibration curve
Empty 550
For creation of customer calibration curve
reference
! Only for checking the instrument !
6.0
LIST OF CALIBRATION CURVES
By pressing one of the arrow keys in the measuring
window for approx. 3 seconds a list of all available
calibration curves opens. Now you can select your
desired curve using the arrow keys and confirm by
pressing .
Electromatic Equipment Co., Inc
-13-
17.0 COMMON REASONS FOR INCORRECT MEASUREMENTS
•
Wrong calibration curve
Double-check the correct selection of the calibration curve before measuring.
The calibration curve „reference“ must NOT be used.
•
Inadequate thickness of the paper pile
A single sheet of paper must NOT be measured. Make sure that the paper pile
below the steel clamps has a thickness of at least 100 mm.
•
Electro conductive material
Every metallic object as well as electro conductive packing material e.g. paper
coloured by soot or wire has a negative effect on the accuracy of your
measurement. Make sure that in an area of 20 cm below the sensor steel
clamps there is no such material.
•
Product temperature out of application range
The application range is between 0 and +60°C.
•
Discrepancy in temperature between device and material
Please ensure that the device and the material under test are being stored at
nearly the same temperature before measuring. A high temperature difference
has a negative effect on the stability of the measurement results.
•
Measurement at the front side
A measurement at the front side of a paper roll will bring different measuring
results. For that you have to select a different calibration curve.
•
Bended steel clamps
If the steel clamps (measuring area) are bended you might get wrong measuring
results.
Please note:
•
After measuring at the running paper roll for a longer period the steel clamps will
get hot – don´t touch! => risk of getting burned!
•
Measuring fine paper at the running roll may cause pressure marks on the paper
=> for possible damages at the paper we refuse any liability.